calo2310 量热原理 -大昌华嘉,激光粒度仪,接触角测量仪,表面/界面张力仪,旋光仪,密度仪,全自动进样器,折光仪,全自动氨基酸分析仪,Zeta电位及纳米粒度仪,比表面和孔隙度分析仪,薄层色谱扫描仪,总有机碳(TOC)分析仪,元素分析仪,水分活度仪,火焰光度计/氯离子分析

大昌华嘉咨询热线:400-821-0778

您现在的位置:首页 >服务 >calo2310 量热原理

calo2310 量热原理 发布时间:2014/05/27 13:43:33

Introduction Sometimes I am being asked: “What is the difference between the Calo 2000 principle and an RC1, an HELSimular, or even our own calorimeters SC1 ... SC4?” The answer is: “The Calo 2000 principle is the first reaction calorimeter commercially available, which allows for the recording of chemical power in absolute Watt and which exhibits practically no time error - except the filter time constant of 60 sec. for the evaluation of the measurement.” To achieve this the baseline has to be set to 0 W exactly. Due to various interferences (temperature ramp, dosing etc.) this is not always possible

点击索取

您索取的服务为: